Skip to main content
Revvity logo
Contact us
Revvity Sites Globally

Select your location.

*e-commerce not available for this region.

australia.webp Australia
austria.webp Austria
belgium.webp Belgium
brazil.webp Brazil *
canada.webp Canada
china.webp China *
denmark.webp Denmark
finland.webp Finland
france.webp France
germany.webp Germany
hong-kong.webp Hong Kong (China) *
india.webp India *
ireland.webp Ireland
italy.webp Italy
japan.webp Japan *
luxembourg.webp Luxembourg
mexico.webp Mexico *
netherlands.webp Netherlands
norway.webp Norway
philippines.webp Philippines *
republic of korea.webp Republic of Korea *
singapore.webp Singapore *
spain.webp Spain
sweden.webp Sweden
switzerland.webp Switzerland
thailand.webp Thailand *
uk.webp United Kingdom
usa.webp United States
Breadcrumb
...
  • ホーム
  • Resources
  • Knowledge Base
  • Application Support Knowledge base (ASK)
  • pHSense
  • pHSense reader settings
pHSense

pHSense reader settings

Section
pHSense reader settings
Critical pHSense assay parameters
pHSense Eu Fab Anti-IgG affinities
pHSense basics
pHSense data reduction and analysis
pHSense probe options
pHSense protocol
pHSense reader settings

Introduction

As explained in the “pHSense Basics” section, it is important to use the correct reader settings to benefit from the full potential and best sensitivity of the pHSense™ dyes, and pHSense signal must be acquired using a compatible plate reader.

All HTRF-certified readers are compatible with pHSense assays; however, TRF reader settings for pHSense assays differ from standard HTRF settings. For optimal assay performance, a delay time of 1,000 µs and an integration (reading) time of 400 µs are typically required. Using standard HTRF settings would miss the benefit from the lifetime sensitivity to pH and may result in a significantly reduced assay window.

The pHSense assay operates in a single detection mode at 620 nm. If a dual-emission mirror is used, the 665 nm signal should be ignored.

For reader configuration, please refer to the documents below. For other plate readers, please contact your local technical support to obtain appropriate pHSense setup recommendations.

BMG

PHERAstar FS View document here
PHERAstar FSX
PHERAstar PLUS

Revvity

EnVision Nexus View document here
EnVision
VICTOR Nivo

The TRF signal evolves over time according to the internalization kinetics of the binder or receptor and the assay conditions (e.g., incubation time, incubation temperature, cell type, culture medium).

Plates can be read multiple times over the course of the incubation period, typically up to 24 hours, without compromising data quality. Please refer to the “Critical pHSense Assay Parameters” section regarding the potential sensitivity to UV illumination of some cell types on some plate readers (also detailed and exemplified in the “Guide ADC binder internalization”.)

For research use only. Not for use in diagnostic procedures.

The information provided above is solely for informational and research purposes only. Revvity assumes no liability or responsibility for any injuries, losses, or damages resulting from the use or misuse of the provided information, and Revvity assumes no liability for any outcomes resulting from the use or misuse of any recommendations. The information is provided on an "as is" basis without warranties of any kind. Users are responsible for determining the suitability of any recommendations for the user's particular research. Any recommendations provided by Revvity should not be considered a substitute for a user's own professional judgment.